Aomori Olympus Co., Ltd.: Medical Device Recall in 2021 - (Recall #: Z-0987-2021)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2021.

Data Source: FDA.
Product Description:

SD-230U-20 ELECTROSURGICAL SNARE, model no. SD-230U-20 - Product Usage: Used with endoscopes to electrosurgically resect tissue within the digestive tract.

Product Classification:

Class II

Date Initiated: January 4, 2021
Date Posted: February 17, 2021
Recall Number: Z-0987-2021
Event ID: 87074
Reason for Recall:

Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.

Status: Terminated
Product Quantity: 7,400,000 (globally); 260,395 (US)
Code Information:

97V, 98V, 9XV, 9YV, 03V, 04V, 05V, 06V, 07V

Distribution Pattern:

US Nationwide distribution.

Voluntary or Mandated:

Voluntary: Firm initiated