Applied Medical Resources Corp: Medical Device Recall in 2013 - (Recall #: Z-0644-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

Laparoscope, general; plastic surgery Used for general, abdominal, gynecological and thoracic minimally invasive surgical procedures to establish a path of entry or to gain access through tissue planes and/or potential spaces for endoscopic instruments.

Product Classification:

Class II

Date Initiated: November 5, 2012
Date Posted: January 16, 2013
Recall Number: Z-0644-2013
Event ID: 63842
Reason for Recall:

Applied Medical is conducting a voluntary recall on a specific lot number of its 12x150mm Kii shielded bladed access systems. This is due to a potential for the blade to remain exposed after shield deployment. Further use of any affected product should cease immediately. The affected model number is CTB71, and the affected lot is 1179107.

Status: Terminated
Product Quantity: 72 units
Code Information:

Lot Number: 1179107

Distribution Pattern:

Worldwide Distribution - USA (nationwide) and France.

Voluntary or Mandated:

Voluntary: Firm initiated