Applied Medical Resources Corp: Medical Device Recall in 2020 - (Recall #: Z-2158-2020)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

Epix latis GRASPER, 5 mm x 35 cm, REF C4130, Qty: 10, Sterile R, CE, Rx ONLY, UDI: (01)00607915110147

Product Classification:

Class II

Date Initiated: March 17, 2020
Date Posted: June 3, 2020
Recall Number: Z-2158-2020
Event ID: 85436
Reason for Recall:

There may be a slightly protruding rivet on the distal end of the laparoscopic grasper which has the potential for tissue to catch on it during use and lead to tissue damage.

Status: Terminated
Product Quantity: 32,760 units
Code Information:

Batch: 1364578/Expiration Date: July 28, 2022; Batch: 1366453/Expiration Date: August 13, 2022; Batch: 1371249/Expiration Date: October 3, 2022; Batch: 1373050/Expiration Date: October 23, 2022; Batch: 1373248/Expiration Date: November 4, 2022; Batch: 1373249/Expiration Date: November 11, 2022; Batch: 1373255/Expiration Date: November 18, 2022; Batch: 1373260/Expiration Date: November 19, 2022; Batch; 1373261/Expiration Date: November 24, 2022; Batch: 1373262/Expiration Date: December 2, 2022; Batch: 1373263/Expiration Date: December 5, 2022; Batch: 1377363/Expiration Date: December 11, 2022; Batch: 1377364/Expiration Date: December 17, 2022

Distribution Pattern:

US: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY OUS: Argentina, Australia, Austria, Belgium, Canada, Chile, Costa Rica, Denmark, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Latvia, Liechtenstein, Luxembourg, Netherlands, New Zealand, Peru, Portugal, Serbia, Singapore, South Africa, Spain, Thailand, United Kingdom

Voluntary or Mandated:

Voluntary: Firm initiated