Applied Medical Resources Corp: Medical Device Recall in 2020 - (Recall #: Z-2158-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
Epix latis GRASPER, 5 mm x 35 cm, REF C4130, Qty: 10, Sterile R, CE, Rx ONLY, UDI: (01)00607915110147
Class II
There may be a slightly protruding rivet on the distal end of the laparoscopic grasper which has the potential for tissue to catch on it during use and lead to tissue damage.
Batch: 1364578/Expiration Date: July 28, 2022; Batch: 1366453/Expiration Date: August 13, 2022; Batch: 1371249/Expiration Date: October 3, 2022; Batch: 1373050/Expiration Date: October 23, 2022; Batch: 1373248/Expiration Date: November 4, 2022; Batch: 1373249/Expiration Date: November 11, 2022; Batch: 1373255/Expiration Date: November 18, 2022; Batch: 1373260/Expiration Date: November 19, 2022; Batch; 1373261/Expiration Date: November 24, 2022; Batch: 1373262/Expiration Date: December 2, 2022; Batch: 1373263/Expiration Date: December 5, 2022; Batch: 1377363/Expiration Date: December 11, 2022; Batch: 1377364/Expiration Date: December 17, 2022
US: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY OUS: Argentina, Australia, Austria, Belgium, Canada, Chile, Costa Rica, Denmark, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Latvia, Liechtenstein, Luxembourg, Netherlands, New Zealand, Peru, Portugal, Serbia, Singapore, South Africa, Spain, Thailand, United Kingdom
Voluntary: Firm initiated