Applied Medical Resources Corp: Medical Device Recall in 2020 - (Recall #: Z-2160-2020)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

DR. THAMES - LAP CHOLE, REF: K0880, UDI: (01)30607915139620

Product Classification:

Class II

Date Initiated: March 17, 2020
Date Posted: June 3, 2020
Recall Number: Z-2160-2020
Event ID: 85436
Reason for Recall:

There may be a slightly protruding rivet on the distal end of the laparoscopic grasper which has the potential for tissue to catch on it during use and lead to tissue damage.

Status: Terminated
Product Quantity: 21 units
Code Information:

Batch: 1378830/Expiration Date: September 8, 2022; Batch: 1379253/Expiration Date: August 13, 2022; Batch: 1380094/Expiration Date: August 22, 2022; Batch: 1381695/Expiration Date: October 30, 2022

Distribution Pattern:

US: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY OUS: Argentina, Australia, Austria, Belgium, Canada, Chile, Costa Rica, Denmark, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Latvia, Liechtenstein, Luxembourg, Netherlands, New Zealand, Peru, Portugal, Serbia, Singapore, South Africa, Spain, Thailand, United Kingdom

Voluntary or Mandated:

Voluntary: Firm initiated