Applied Medical Technology Inc: Medical Device Recall in 2016 - (Recall #: Z-1497-2016)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2016.
Data Source: FDA.
Product Description:
AMT G-Jet¿ Low-Profile Transgastric-Jejunal Feeding Device 14F 1.5cm 45cm SINGLE USE ONLY Rx ONLY STERILE. For use in patients who cannot absorb adequate nutrition through the stomach.
Product Classification:
Class II
Date Initiated: January 26, 2016
Date Posted: April 20, 2016
Recall Number: Z-1497-2016
Event ID: 73339
Reason for Recall:
The firm received complaints that the product is shorter in length than indicated on the label.
Status: Terminated
Product Quantity: 20 units
Code Information:
Model #GJ-1415-45, Lot # 15104750, Exp Date: 2018/09
Distribution Pattern:
Distributed in the states of AZ, CA, DE, IN, MI, NM, PA & VA.
Voluntary or Mandated:
Voluntary: Firm initiated