Product Description:

BONE MARROW ASPIRATION NEEDLE 15G x 2.688 MAX and 15G x 4 MAX Intended for the purpose of harvesting bone and/or bone marrow specimens

Product Classification:

Class II

Date Initiated: January 18, 2017
Date Posted: February 22, 2017
Recall Number: Z-1194-2017
Event ID: 76292
Reason for Recall:

The recalling firm received a complaint from one of their customers of a potential defect in the forming of the primary package of the blister that may compromise the sterility of the product inside the tray.

Status: Terminated
Product Quantity: 497 boxes
Code Information:

lots: 11156986, 11158618, 11158620

Distribution Pattern:

Worldwide Distribution - US (nationwide) and Internationally to Switzerland, Canada, Australia, New Zealand, Ecuador, Colombia, and Japan.

Voluntary or Mandated:

Voluntary: Firm initiated