Argon Medical Devices, Inc: Medical Device Recall in 2021 - (Recall #: Z-1076-2021)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2021.

Data Source: FDA.
Product Description:

Argon Medical Devices TLAB, Transjugular Liver Biopsy System, REF: TL - 18C, 1/18 GA Flexcore Biopsy Needle, 1/7F Introducer Sheath W/Curved Metal Stiffener, 4/Tissue Removal Swabs, 1/5f straight catheter, 1/5F Curved Catheter Sterile EO,

Product Classification:

Class II

Date Initiated: December 10, 2020
Date Posted: February 17, 2021
Recall Number: Z-1076-2021
Event ID: 87000
Reason for Recall:

Due to a manufacturing error their is a potential that the sheath within the biopsy kit can separate during use if the tip is flexed or experiences bending stresses. This could result in need for surgical intervention to preclude or reverse permanent damage to a body structure or permanent impairment of a body function.

Status: Terminated
Product Quantity: 1759 units
Code Information:

Lot Number: 1394204, 1424962, 1464709, 1473181, 1477229, 1481735, 1495182

Distribution Pattern:

US: AL AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA,, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX,, UT, VA, WA, WI, WV, WY OUS: Switzerland, Saudi Arabia, Hong Kong, Thailand, Portugal, Taiwan, Austria, Brazil, Netherlands, Italy, Germany, Colombia, Singapore, Canada, View Nam, Slovakia, Spain

Voluntary or Mandated:

Voluntary: Firm initiated