Arjohuntleigh Magog: Medical Device Recall in 2018 - (Recall #: Z-2660-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

Quick Connect 272 KG Scale used with Maxi Sky 2 ceiling lift The Scale is intended to be used with patient lifters to measure the weight of a patient who cannot stand on a scale. It is designed to be used indoors in hospitals, nursing homes, home care or health care facilities.

Product Classification:

Class II

Date Initiated: February 12, 2018
Date Posted: August 15, 2018
Recall Number: Z-2660-2018
Event ID: 80465
Reason for Recall:

The spreader bar can detach from the scale attachment under certain conditions.

Status: Terminated
Product Quantity: 874 Units (both styles)
Code Information:

Product Code 700-19485: starting from serial number 300146551 up to serial number 300220716.

Distribution Pattern:

Worldwide distribution: US (nationwide) to states of: CO, FL, IA, IL, KY, MN, NY, PH, and WA; and countries of: Australia, Belgium, Brazil, Canada, China, France, Germany, Hong Kong, Ireland, Italy, Japan, Netherlands, Norway, Singapore, Swiss, Turkey, United Kingdom, and United Arab Emirates.

Voluntary or Mandated:

Voluntary: Firm initiated