ARJOHUNTLEIGH POLSKA Sp. z.o.o.: Medical Device Recall in 2016 - (Recall #: Z-2718-2016)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2016.

Data Source: FDA.
Product Description:

ArjoHuntleigh Sara Combilizer; Intended to facilitate early immobilization, rehabilitation, and care of patients.

Product Classification:

Class II

Date Initiated: June 29, 2016
Date Posted: September 14, 2016
Recall Number: Z-2718-2016
Event ID: 74506
Reason for Recall:

ArjoHuntleigh has concluded that a device tipping hazard may occur on the current Sara Combilizer lifting mechanism, providing a potential for risk to the patient or the caregiver. The tipping phenomenon has been found during internal device testing. To date, there have been no adverse events or customer complaints related to the tipping phenomenon. The tipping hazard may occur when fully loaded Sara Combilizer in "stretcher mode" is lifted to the highest position (approximately 984 mm = 38 3/4") and tilted backwards to the angle of 25 degrees¿-28 degrees¿. Although such scenario has been defined as highly unlikely to occur during use with patients, as a commitment to quality, ArjoHuntleigh is providing customers with information on action needed. The Sara Combilizer may stay in use until the upgrade will be performed, on the condition that the tilt angle in the device ''stretcher mode" will be no greater than minus 15 degrees¿ and the device is not loaded over 330 lbs (150 kg).

Status: Terminated
Product Quantity: 106
Code Information:

Device Listing Number E101987, Model Number APA1000-XX; Serial numbers P0359671 , P0360103 , P0338345 , P0354548 , P0357690 , P0359672 , P0359673 , P0348548 , P0349629 , P0361336 , P0361337 , P0361875 , P0361876 , P0361877 , P0360104 , P0360105 , P0346000 , P0373592 , P0357691 , P0349634 , P0354549 , P0349627 , P0339257 , P0321458 , P0299662 , P0325930 , P0325931 , P0325933 , P0310854 , P0344175 , P0356357 , P0356362 , P0356363 , P0360990 , P0375493 , P0298175 , P0309635 , P0310848 , P0323622 , P0323624 , P0325167 , 1404000118 , P0296131 , P0296315 , P0298465 , P0319583 , P0333285 , P0333288 , P0339264 , P0343583 , P0343584 , P0343585 , P0343684 , P0344174 , P0353246 , P0364628 , 1404000112 , P0333284 , P0333290 , P0294982 , P0294983 , P0295762 , P0298158 , P0309472 , P0317162 , P0325970 , P0331310 , 1404000113 , 1404000119 , 1404000120 , P0325967 , P0332385 , P0333283 , P0339574 , P0296130 , P0296318 , P0321460 , P0332383 , 1404000115 , P0331312 , P0331311 , P0331318 , P0331319 , P0325968 , P0295846 , P0294981 , P0294985 , P0295847 , P0332386 , P0294984 , P0332384 , 1404000116 , 1404000117 , P0321461 , 1404000114 , P0294986 , P0298176 , P0323011 , P0361422 , P0361423 , P0363456 , P0357692 , P0360102 , P0328718 , P0328720 , P0328724 .

Distribution Pattern:

CO, AL, MA, NC, MI, NY, FL, MD, CA, VA, CT, TX, and IL; Australia, Austria, Denmark, Germany, Israel, Japan, Norway, Sweden, Switzerland, UAE, and UK.

Voluntary or Mandated:

Voluntary: Firm initiated