Arkray Factory USA, Inc.: Medical Device Recall in 2016 - (Recall #: Z-0631-2016)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2016.

Data Source: FDA.
Product Description:

arkray SPOTCHEM II Basic Panel - 1 Reagent test strips, Product # 11726, MEDTEST DX Catalog # 77484. Calcium, Blood Urea Nitrogen, Glucose, Albumin, Creatinine. For the in vitro quantitative determination of calcium, blood urea nitrogen (BUN), glucose, albumin, and creatinine.

Product Classification:

Class I

Date Initiated: December 18, 2015
Date Posted: January 27, 2016
Recall Number: Z-0631-2016
Event ID: 72860
Reason for Recall:

One lot each of SPOTCHEM II Basic PANEL-1 Reagent Test Strip and SPOTCHEM II Glucose Reagent Test Strip are being recalled due to a lowered reaction rate in the reagent pad for glucose. The affected lots will display a result of approximately 300 to 400 mg/dL for any glucose value that exceeds the upper limit of 416 mg/dL.

Status: Terminated
Product Quantity: 11 boxes (25 foil packaged test strips per box)
Code Information:

Lot Number PN5C26

Distribution Pattern:

Distributed in the states of FL, IL, KY, MI, NC, NY, OH, and TN.

Voluntary or Mandated:

Voluntary: Firm initiated