AROA Biosurgery, LTD.: Medical Device Recall in 2018 - (Recall #: Z-0435-2019)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2018.
Data Source: FDA.
Product Description:
OviTex 2S Reinforced BioScaffold 10x12cm, Part Number F10258-1012G
Product Classification:
Class II
Date Initiated: April 6, 2018
Date Posted: November 21, 2018
Recall Number: Z-0435-2019
Event ID: 81166
Reason for Recall:
Degradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no longer meeting the pre-defined ball burst specification respective to the number of tissue layers.
Status: Terminated
Product Quantity: 881 total
Code Information:
ERT-6H16 ERT-6K01 ERT-7J06 ERT-7L15
Distribution Pattern:
The products were distributed to the following US states: AL, CA, FL, IN, MA, MI, NH, and NY.
Voluntary or Mandated:
Voluntary: Firm initiated