Arrow International Inc: Medical Device Recall in 2013 - (Recall #: Z-1057-2013)
See the recall detail below. You can also see other recalls from the same firm in 2013.
Arrow/Teleflex Medical Berman Angiographic Balloon Catheter and Reverse Berman Angiographic Balloon Catheter. The Arrow Berman(TM) Angiographic Catheter is indicated for use in right heart catheterization for cardiac angiography, and measurement of the right heart and pulmonary artery pressures. It can be sued to inject radiopaque media and measure pressures in any chamber or vessel that can be entered from the right heart. The Arrow Berman(TM) Angiographic Catheter is also indicated for use in Balloon Occlusion Femoral Angiography (BOFA) of the ipsilateral leg for adult patients. The Arrow Reverse Berman(TM) Angiographic Catheter is indicated for balloon occlusion pulmonary angiography.
Class II
The product specifications sheet, part of the Instructions for Use (IFU), for the specified product numbers, reference Renografin-76 contrast medium (viscosity of 8.4 centipoise), which is no longer available. Firm is in the process of updating the labeling for these products. In the meantime, when choosing an alternative contrast medium, health care providers are instructed to select a contrast
Part Numbers --AI-07035, AI-07131, AI-07135, AI-07037, AI-07132, AI-07136, AI-07130, AI-07133, AI-07137, AI-07130-J, AI-07134, AND AI-07138.
Worldwide Distribution - USA including Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI and WV. and the countries of Australia, Dominican Republic, Uruguay, Ecuador, Chile, Taiwan, Pakistan, Colombia, Argentina, Korea Republic, New Zealand, Guatemala, Thailand, Peru, Bahamas, Philippines, Trinidad and Tobago, Japan, Ireland, India, Singapore, China, Canada and Mexico.
Voluntary: Firm initiated