Arrow International Inc: Medical Device Recall in 2017 - (Recall #: Z-2698-2017)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2017.
Data Source: FDA.
Product Description:
Arrow EVOLUTION Pressure Injectable PICC with Chlorag+ard Antimicrobial and Antithrombogenic Technology
Product Classification:
Class II
Date Initiated: June 12, 2017
Date Posted: July 19, 2017
Recall Number: Z-2698-2017
Event ID: 77604
Reason for Recall:
Arrow is recalling additional lots that were identified as part of an active recall. Arrow is recalling the affected product due to the possibility that the catheter peel-away component hub tabs may prematurely detach when the practitioner begins to peel apart the sheath body from the catheter.
Status: Terminated
Product Quantity: 9,037 units in the U.S. and 4,500 units internationally
Code Information:
Device Listing # D149129 Material # ASK-45552-RWJ4 PR-45552-HPHNL
Distribution Pattern:
Distributed to SC, AL, NJ, IN, MA, GA, CA, PA, AZ, VA , WA and Bangkok
Voluntary or Mandated:
Voluntary: Firm initiated