Arrow International Inc: Medical Device Recall in 2019 - (Recall #: Z-0941-2019)
See the recall detail below. You can also see other recalls from the same firm in 2019.
AF w/AS 550x 0.5 7 ml/hr, 5ml 30 LO CPNB, Product code MVAX30XL CPNB, Arrow AutoFuser Disposable Pain Control Pump
Class II
Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.
74F1700085 74F1703357
Nationwide distribution.
Voluntary: Firm initiated