Arrow International Inc: Medical Device Recall in 2019 - (Recall #: Z-0943-2019)
See the recall detail below. You can also see other recalls from the same firm in 2019.
AF550x5ML/H 2ML 60LO 6" DUAL CATHS, 4.5", Product MT5060XLYK5 CP, Arrow AutoFuser Disposable Pain Control Pump
Class II
Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.
74F1700794 74J1700676 74J1700190
Nationwide distribution.
Voluntary: Firm initiated