Arrow International Inc: Medical Device Recall in 2019 - (Recall #: Z-0948-2019)
See the recall detail below. You can also see other recalls from the same firm in 2019.
AF 550 x 5ML/HR 2ML 60LO CPNB, Product code MT5060XL CPNB, Arrow AutoFuser Disposable Pain Control Pump
Class II
Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.
74A1700874 74G1702797 74F1700146 74H1701106 74F1701964
Nationwide distribution.
Voluntary: Firm initiated