Arrow International Inc: Medical Device Recall in 2019 - (Recall #: Z-0949-2019)
See the recall detail below. You can also see other recalls from the same firm in 2019.
AF 550 x 5ML/HR CPNB, Product code MC0050XL CPNB, Arrow AutoFuser Disposable Pain Control Pump
Class II
Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.
74C1601434 74E1602096 74C1602395 74F1602444 74C1703028 74F1700186 74D1700428 74G1702048 74D1701032
Nationwide distribution.
Voluntary: Firm initiated