Arrow International Inc: Medical Device Recall in 2019 - (Recall #: Z-0955-2019)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2019.
Data Source: FDA.
Product Description:
AF w/AS 550 x 1-14ML/HR 2ML-30LO CPNB, Product code MVBT30XL CPNB, Arrow AutoFuser Disposable Pain Control Pump
Product Classification:
Class II
Date Initiated: October 2, 2018
Date Posted: March 6, 2019
Recall Number: Z-0955-2019
Event ID: 81375
Reason for Recall:
Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.
Status: Terminated
Product Quantity: 1
Code Information:
74H1701904
Distribution Pattern:
Nationwide distribution.
Voluntary or Mandated:
Voluntary: Firm initiated