Arrow International Inc: Medical Device Recall in 2020 - (Recall #: Z-2409-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
Arrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 SPANISH, Product Code IAP-0400E
Class I
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
Serial Number Ranges: 150301V 160119V 160429V through 160430V 160634V 160643V
Worldwide distribution.
Voluntary: Firm initiated