Arrow International Inc: Medical Device Recall in 2020 - (Recall #: Z-2412-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
Arrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE ITALIAN, Product Code IAP-0500I
Class I
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
Serial Number Ranges: AUTOCAT2 WAVE ITALIAN IAP-0500I 160516W 160619W 160804W 160828W 170209W through 170210W 170302W 180309W
Worldwide distribution.
Voluntary: Firm initiated