Arrow International Inc: Medical Device Recall in 2020 - (Recall #: Z-2415-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
Arrow AutoCAT2 Intra-Aortic Balloon Pump AEROAUTOCAT 2 WAVE, Product Code IAP-0535
Class I
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
Serial Number Range: AEROAUTOCAT 2 WAVE IAP-0535 150907W 151030W 151101W 151104W 151106W 160305W through 160306W 160743W 160824W 190119W through 190120W 190221W 190223W through 190226W 191047W
Worldwide distribution.
Voluntary: Firm initiated