Asante Solutions, Inc.: Medical Device Recall in 2014 - (Recall #: Z-0596-2014)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2014.
Data Source: FDA.
Product Description:
Asante Comfort Infusion Set with 110cm tubing & 13mm cannula packaged in 5-packs,singles, or inside a Combination Kit (FG-5143-4). Manufactured for: Asante Solutions Sunnyvale, CA; Manufactured by: Unomedical Lejre, Denmark
Product Classification:
Class II
Date Initiated: November 19, 2013
Date Posted: January 8, 2014
Recall Number: Z-0596-2014
Event ID: 66895
Reason for Recall:
Reports of the infusion tubing disconnecting from the connector, resulting in leakage of insulin from the infusion set and providing patient with less insulin than intended.
Status: Terminated
Product Quantity: 635 kits
Code Information:
Asante Catalog number: FG-4143-5, FG-4143T. Unomedical Item number: 87-110-2562 Lot numbers: 0230120; 0230124 5031108.
Distribution Pattern:
Nationwide Distribution.
Voluntary or Mandated:
Voluntary: Firm initiated