Asante Solutions, Inc.: Medical Device Recall in 2014 - (Recall #: Z-0597-2014)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2014.
Data Source: FDA.
Product Description:
Asante Comfort Infusion Set with 110cm tubing & 17mm cannula packaged in 5-packs, singles, or inside a Combination Kit (FG-5147-4). Manufactured for: Asante Solutions Sunnyvale, CA; Manufactured by: Unomedical, Lejre, Denmark
Product Classification:
Class II
Date Initiated: November 19, 2013
Date Posted: January 8, 2014
Recall Number: Z-0597-2014
Event ID: 66895
Reason for Recall:
Reports of the infusion tubing disconnecting from the connector, resulting in leakage of insulin from the infusion set and providing patient with less insulin than intended.
Status: Terminated
Product Quantity: 352
Code Information:
Asante Catalog number:FG-4147-5, FG-4147T Unomedical Item number:87-110-2662; Lot numbers: 0230121 0230125 5025227.
Distribution Pattern:
Nationwide Distribution.
Voluntary or Mandated:
Voluntary: Firm initiated