Asante Solutions, Inc.: Medical Device Recall in 2014 - (Recall #: Z-0598-2014)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2014.
Data Source: FDA.
Product Description:
Asante Conset Infusion Set with 60cm tubing & 6mm cannula packaged in 5-packs,singles,or inside a Combination Kit (FG-5226-4) Distributed by Asante Solutions, Sunnyvale, CA; Manufactured by Convatec; Lejre, Denmark.
Product Classification:
Class II
Date Initiated: November 19, 2013
Date Posted: January 8, 2014
Recall Number: Z-0598-2014
Event ID: 66895
Reason for Recall:
Reports of the infusion tubing disconnecting from the connector, resulting in leakage of insulin from the infusion set and providing patient with less insulin than intended.
Status: Terminated
Product Quantity: 1392 kits
Code Information:
Asante Catalog number: FG-4226-5, G-4147T; Unomedical Item number:43-060-5266; Lot numbers: 0230126 5005801 5030167.
Distribution Pattern:
Nationwide Distribution.
Voluntary or Mandated:
Voluntary: Firm initiated