Aspen Surgical Products, Inc.: Medical Device Recall in 2021 - (Recall #: Z-2388-2021)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2021.
Data Source: FDA.
Product Description:
Jessie Sterile Arm Suspension Kit, Product Code 711300, (previous product code 100401), Sample Kit S711300
Product Classification:
Class II
Date Initiated: July 23, 2021
Date Posted: September 8, 2021
Recall Number: Z-2388-2021
Event ID: 88476
Reason for Recall:
Affected products were manufactured with materials found to have a bioburden level exceeding the limit at which the product's sterilization process was validated. As a result, the firm cannot provide full assurance that the affected products were effectively sterilized, and the product may compromise a sterile field and increase the risk of a surgical site infection.
Status: Terminated
Product Quantity: 1429
Code Information:
All lots manufactured between July 9, 2019 to July 9, 2021
Distribution Pattern:
US nationwide distribution. No foreign distribution.
Voluntary or Mandated:
Voluntary: Firm initiated