AtriCure, Inc.: Medical Device Recall in 2019 - (Recall #: Z-0830-2019)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2019.
Data Source: FDA.
Product Description:
COBRA FUSION 150 Ablation System, Catalog # (REF) 700-001
Product Classification:
Class II
Date Initiated: January 22, 2019
Date Posted: February 20, 2019
Recall Number: Z-0830-2019
Event ID: 81932
Reason for Recall:
There is a potential for patients undergoing ablation to experience an elevated risk for TE; during stand-alone off-pump procedures where cardio pulmonary bypass (CPB) was not utilized.
Status: Terminated
Product Quantity: 6176 units
Code Information:
UDI - none Lot Numbers - All lots within expiry
Distribution Pattern:
AL, AR, AZ, CA, CT, CO, FL, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA & WI Internationally - Germany, Italy, Netherlands, France, Austria, United Kingdom, Poland, Czech Republic, United Arab Emirates, Sweden, Switzerland, Kuwait, Belgium and Latvia
Voluntary or Mandated:
Voluntary: Firm initiated