AURIS HEALTH INC: Medical Device Recall in 2024 - (Recall #: Z-0051-2025)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

Monarch Platform (Endoscopy), REF: MON-000005-01, MON-000005-01R. Monarch Platform (Bronchoscopy), REF: MON-000006, MON-000006-RFB.

Product Classification:

Class II

Date Initiated: August 2, 2024
Date Posted: October 16, 2024
Recall Number: Z-0051-2025
Event ID: 95336
Reason for Recall:

Endoscopy/bronchoscopy/urology devices may have a strain relief defect with some cart and tower power cords that may cause an electrical short, which may lead to system shutdown, which can lead to prolonged procedure or case conversion, and if the user (e.g., operating room staff) touches exposed wires or conductive fluids in contact with the exposed wires, there is a chance of electrical shock.

Status: Ongoing
Product Quantity: 176
Code Information:

Cart Power Cord 20' 14AWG BLK, PN 304-003613-00. Tower Power Cord 20' 14AWG BLK, PN 304-003607-00. REF/UDI-DI/Serial Numbers: MON-000005-01/+B634MON000005010/110024, 13, 110014, 110016, 110040, 110038, 110027, 110035, 110022, 110034, 1806P2, 120170, 110045, 110039, 110049, 110047, 110052, 110030, 110032, 110021, 110043, 110050, 110029, 110033, 110046, 110048, 110018, 110015, 110051, 1806P1, 110012, 14, 110041, 110036; MON-000005-01R/+B634MON00000501R0/110020, 110017, 110031, 110028, 120042, 110013. MON-000006/+B634MON0000060/120088, 120057, 120046, 120058, 120121, 120152, 120015, 120160, 120068, 120167, 120025, 120030, 120044, 120093, 120040, 120063, 120092, 120173, 120036, 120064, 120122, 120056, 120128, 120017, 120089, 120075, 120031, 120169, 120113, 120168, 120055, 120148, 120014, 120060, 120124, 120164, 120123, 120007, 120006, 120004, 120120, 120126, 120082, 120047, 120051, 120020, 120103, 120153, 120108, 120037, 120074, 120005, 120130, 120095, 120096, 120009, 120111, 120022, 120054, 120163, 120027, 120090, 120141, 120072, 120155, 120101, 120143, 120125, 120052, 120100, 120050, 120098, 120045, 120131, 120013, 120105, 120165, 120136, 120029, 120041, 120139, 120137, 120053, 120028, 120034, 120085, 120019, 120049, 120209, 120144, 120161, 120001, 120145, 120127, 120011, 120070, 120076, 120134, 120117, 120077, 120149, 120102, 120002, 120010, 120003, 120023, 120146, 120043, 120071, 120107, 120065, 120116, 120150, 120066, 120166, 120033, 120079, 120008, 120132, 120086, 120032, 120073, 120016, 120080, 120078, 120012, 120084, 120083, 120038, 120067, 120021; MON-000006-RFB/+B634MON0000060/120081, 120026, 120091, 120106, 120104.

Distribution Pattern:

Worldwide - US Nationwide distribution including in the states of MN, FL, NC, AK, NJ, NY, VA, WI, AZ, MS, LA, MI, MA, MT, WV, GA, DE, TX, CA, OH, IN, SC, PA, CT, ND, MO, DC, IL, NE, ID, WA, TN, OR, AL, SD, CO, KY, NM, VT, MD and the countries of Canada.

Voluntary or Mandated:

Voluntary: Firm initiated