Autochair: Medical Device Recall in 2018 - (Recall #: Z-1420-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

Product: Milford Smartbase Model # 9934105 Pride SmartBase Pride product codes: LSMPLCG A & LSMPLCG B

Product Classification:

Class II

Date Initiated: March 7, 2018
Date Posted: April 25, 2018
Recall Number: Z-1420-2018
Event ID: 79627
Reason for Recall:

Recalling firm was informed that as a result of combination of a weak spring and incorrect alignment of the locking pin for latching the leg positions, product legs folded inwards while lifting a person off a bed and the individual was dropped.

Status: Ongoing
Product Quantity:
Code Information:

All Lots

Distribution Pattern:

PA, and worldwide

Voluntary or Mandated:

Voluntary: Firm initiated