Product Description:
BALLARD ACCESS Closed Suction System for Neonates/Pediatrics, Y-Manifold.
Product Classification:
Class I
Date Initiated: February 3, 2023
Date Posted: March 22, 2023
Recall Number: Z-1189-2023
Event ID: 91663
Reason for Recall:
Potential for product manifold to crack during use on patients, possibly causing leaks in the respiratory circuit and can lead to the patient receiving inadequate ventilation.
Status: Terminated
Product Quantity: 780 units
Code Information:
All lots are affected. Stock Code (Lot Number): B30511 (20098360), B30611 (20098361), B30711 (20097200), B30811 (20098362), B31011 (20098363).
Distribution Pattern:
US Nationwide distribution in the state of Oklahoma.
Voluntary or Mandated:
Voluntary: Firm initiated