B. Braun Medical, Inc.: Medical Device Recall in 2014 - (Recall #: Z-0921-2014)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2014.
Data Source: FDA.
Product Description:
B. Braun AccuFlo Elastomeric Infusion Pump 200ML/H 100 ML Generally used for delivery of antibiotic and local anesthesia medications.
Product Classification:
Class II
Date Initiated: December 13, 2013
Date Posted: February 12, 2014
Recall Number: Z-0921-2014
Event ID: 67172
Reason for Recall:
Internal testing of two lots of the AccuFlo CT-2000-100 Elastomeric Infusion Pump identified flow rates which are outside the specified infusion time and may result in overly rapid infusion of medications.
Status: Terminated
Product Quantity: 30, 240 units
Code Information:
Catalog #CT-2000-100lots 3H0125F151 and 3H0225F151, expiration 7/2018.
Distribution Pattern:
US Distribution one importer/wholesale distributor in MO.
Voluntary or Mandated:
Voluntary: Firm initiated