B. Braun Medical, Inc.: Medical Device Recall in 2024 - (Recall #: Z-3144-2024)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

AET36 ANESTH EXT SET - Anesthesia Extension Set - Normally Closed Dual Manifold, Four-way Stopcock, Model/Catalog Number: 472036

Product Classification:

Class II

Date Initiated: August 8, 2024
Date Posted: September 25, 2024
Recall Number: Z-3144-2024
Event ID: 95153
Reason for Recall:

Potential for product leakage from the Spin Lock Connector on the distal end of the device due to the migration of excess solvent during the assembly process.

Status: Ongoing
Product Quantity: 19050 units
Code Information:

Model No 472036; UDI-DI 04046964184659; Lots 0061898815, 0061905158, 00VL872690, 00VL882317 and 00VL882319

Distribution Pattern:

US Nationwide distribution.

Voluntary or Mandated:

Voluntary: Firm initiated