Bard Access Systems Inc.: Medical Device Recall in 2021 - (Recall #: Z-1661-2021)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2021.

Data Source: FDA.
Product Description:

REF 9617408, Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet, Full Kit, 4F, Single-Lumen, TLS, UDI: (01)10801741036009 - Product Usage: is designed for use when central venous catheterization is prescribed.

Product Classification:

Class II

Date Initiated: April 7, 2021
Date Posted: May 26, 2021
Recall Number: Z-1661-2021
Event ID: 87594
Reason for Recall:

Catheters contain two issues: 1. the inner diameter of the distal connector may be reduced resulting in inability to pass the catheter through the connector and 2. the catch slot on the distal connecter may be deformed resulting in the inability to properly lock

Status: Terminated
Product Quantity: N/A
Code Information:

Lot No. reep1720 redy0653 REEN1127 REEN1905 reeq0634 rees0249 reet2073 reev1213 reew4243

Distribution Pattern:

Worldwide distribution - US Nationwide distribution in the states of AK, AL, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MA, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI and the countries of Australia, Belgium, Canada, China, Japan, Taiwan, Hong Kong, Malaysia, Chile, Bangladesh, Korea, Brazil, India, Singapore, Thailand.

Voluntary or Mandated:

Voluntary: Firm initiated