Bard Medical Division: Medical Device Recall in 2018 - (Recall #: Z-2536-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

Arctic Sun Temperature Management System, Model 5000. The Arctic Sun¿ 5000 model is a non-invasive, thermal regulating system, indicated for monitoring and controlling patient temperature within a range of 32¿C to 38.5¿C (89.6¿F to 101.3¿F) in adult and pediatric patients of all ages.

Product Classification:

Class II

Date Initiated: June 13, 2018
Date Posted: August 1, 2018
Recall Number: Z-2536-2018
Event ID: 80375
Reason for Recall:

An improper wiring connection on the chiller assembly may result in overheating of the connection and potential failure of certain components in the system.

Status: Terminated
Product Quantity: 5608 Units Total
Code Information:

Model #5000 Product codes: 5000-00-00, 5000-00-00L, 5000-00-00E, 5000-01-01, 5000-01-01L, 5000-01-02, 5000-01-03, 5000-01-04, 5000-01-05, 5000-00-06, 5000-01-07, 5000-01-08. All serial numbers are affected.

Distribution Pattern:

Domestic Distribution: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL,GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. Worldwide Distribution: Canada, Austria, Bahrain, Belgium, Brazil, Brussels China, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Gibraltar, Greece, Guam, Hungary, Iran, Italy, Jordan, Korea, Luxemburg, Malaysia, Netherlands, Norway, Oman, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, UK, Vietnam.

Voluntary or Mandated:

Voluntary: Firm initiated