Bard Peripheral Vascular Inc: Medical Device Recall in 2013 - (Recall #: Z-1533-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

BARD PERIPHERAL VASCULAR VACCESS¿ PTA Balloon Dilatation Catheters, 8mm x 4cm x 80cm. Product Usage: Vaccess¿ PTA Balloon Dilatation Catheters are recommended for use in Percutaneous Transluminal Angioplasty of the femoral, iliac and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also recommended for post-dilatation of balloon expandable stents, self-expanding stents and stent grafts in the peripheral vasculature. This catheter is not for use in coronary arteries. Vaccess¿ PTA Balloon Dilatation Catheters are supplied sterile and intended for single use.

Product Classification:

Class II

Date Initiated: May 28, 2013
Date Posted: June 19, 2013
Recall Number: Z-1533-2013
Event ID: 65395
Reason for Recall:

Bard Peripheral Vascular (BPV) has confirmed that some product code/lot number combination of devices may have the incorrect balloon diameter labeled on the inflation hub. Specifically, the balloon size printed on the hub may read 7mm x 4cm, when the actual balloon size is 8mm x 4cm. All other product labeling is consistent with the actual 8mm x 4cm balloon size.

Status: Terminated
Product Quantity: 415
Code Information:

Product Codes and Lot Numbers, respectively: VA8084R, 93JW0056 VA8084, 937W0090

Distribution Pattern:

US Nationwide Distribution

Voluntary or Mandated:

Voluntary: Firm initiated