Bard Peripheral Vascular Inc: Medical Device Recall in 2025 - (Recall #: Z-0887-2026)
See the recall detail below. You can also see other recalls from the same firm in 2025.
Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T,
Class II
Acute drainage catheters, indicated for use in removing fluid during acute drainage/aspiration procedure, manufactured with safety indicators that erroneously show catheter in sharp position when catheter is in blunt position, which may lead users to believe device is not yet in pleural or peritoneal space when it is, may cause discomfort, delay to treatment, injury to internal organs or vessels.
UDI/Lot/Expiration: (01)10885403108365(17)271130(10)0001623320/0001623320/20271130; (01)10885403108365(17)271130 (10)0001623069/0001623069/20271130
US Nationwide distribution in the states of CA, MN, IN, MI, LA.
Voluntary: Firm initiated