Bard Peripheral Vascular Inc: Medical Device Recall in 2025 - (Recall #: Z-0887-2026)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2025.

Data Source: FDA.
Product Description:

Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T,

Product Classification:

Class II

Date Initiated: November 6, 2025
Date Posted: December 17, 2025
Recall Number: Z-0887-2026
Event ID: 97864
Reason for Recall:

Acute drainage catheters, indicated for use in removing fluid during acute drainage/aspiration procedure, manufactured with safety indicators that erroneously show catheter in sharp position when catheter is in blunt position, which may lead users to believe device is not yet in pleural or peritoneal space when it is, may cause discomfort, delay to treatment, injury to internal organs or vessels.

Status: Ongoing
Product Quantity: 1240
Code Information:

UDI/Lot/Expiration: (01)10885403108365(17)271130(10)0001623320/0001623320/20271130; (01)10885403108365(17)271130 (10)0001623069/0001623069/20271130

Distribution Pattern:

US Nationwide distribution in the states of CA, MN, IN, MI, LA.

Voluntary or Mandated:

Voluntary: Firm initiated