Baxter Healthcare Corp.: Medical Device Recall in 2013 - (Recall #: Z-0189-2014)
See the recall detail below. You can also see other recalls from the same firm in 2013.
SIGMA SPECTRUM Infusion Pump with Master Drug Library component (SHARP LCD screen, 55074-2 revision B) Product Usage: Intended to be used as a visual aid during the controlled administration of intravenous fluids. These fluids may include pharmaceutical drugs, blood, blood products and mixtures of required patient therapy.
Class II
Baxter Healthcare Corporation is recalling the LCD display screen on the SIGMA SPECTRUM Infusion Pump with Master Drug Library for not meeting standards for withstanding an electrostatic discharge.
Product Code: 35700BAX & 35700ABB; Serial Numbers: 433713, 435563, 704056, 707290, 715315, 724216, 740127, 752318, 752447, 753675, 754947, 759766, 765934, 770034, 772017, 773470, 776507, 781016, 781969, 783352, 783454, 787567, 789720, 791225, 794828, 794990, 797185, 804798, 808393, 808764, 810971, 812076, 818718, 821213, 824067, 824198, 839143, 850028, 850577, 852007, 852431, 855450, 860201, 864799, 876765, 877548, 877944, 878637, 879810, 879886, 880022, 884707, 884778, 885565, 889416, 889814, 890949, 890987, 894957, 895413, 896662, 896912, 897576, 900401, 904778, 913511, 915035, 915981, 916173, 916948, 919159, 921162, 921937, 922515, 923861, 945583, 953011, 953360, 953995, 955881, 959847, 967694, 978892, 981692, 986022, 997333, 998512, 1002325, 1018432, 1018436, 1018737
Worldwide Distribution - USA Nationwide in the states of: AR, AZ, CA, CO, CT, FL, HI, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NH, NM, NY, OH, OK, OR, PA, TN, UT, VA, WA and WV and the country Canada.
Voluntary: Firm initiated