Baxter Healthcare Corp.: Medical Device Recall in 2016 - (Recall #: Z-0672-2017)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2016.

Data Source: FDA.
Product Description:

V6 Rear Case Assembly, Product Code 35701, a replacement part for the SIGMA Spectrum Infusion Pump -- SIGMA LLC Medina, NY 14103 --- NOTE: The speakers are a part of the rear case assembly located in the back of the infusion pump. The speaker is intended to generate audible tones during each keystroke, and during alarm conditions. However, some speakers have a variation in their internal wiring that increases the potential for the speaker to fail prematurely. Spectrum Infusion Pumps with non-functional speakers will continue to display visual alarm indications, and will otherwise continue to operate as intended. Product Code 35701 is a replacement assembly that is shipped to customers for self-repair of the Spectrum Infusion Pump. It is not a registered finished good.

Product Classification:

Class II

Date Initiated: September 28, 2016
Date Posted: December 7, 2016
Recall Number: Z-0672-2017
Event ID: 75311
Reason for Recall:

A correction was issued for the SIGMA Spectrum Infusion Pump with Master Drug Library and Rear Case Assembly in order to replace speakers installed in pumps or included in spare part rear case assemblies that have an increased potential to fail prematurely.

Status: Terminated
Product Quantity: 439 units
Code Information:

Product Code 35701; distributed between 02/04/2015 - 06/02/2015.

Distribution Pattern:

Nationwide Distribution, Puerto Rico and Canada

Voluntary or Mandated:

Voluntary: Firm initiated