Baxter Healthcare Corporation: Medical Device Recall in 2014 - (Recall #: Z-2126-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

FlowCOUPLER. An implantable device that is used to detect blood flow in vessels. Used in the end-to-end anastomosis of veins and arteries normally encountered in microsurgical and vascular reconstructive procedures. The FlowCOUPLER System is intended to detect blood flow and confirm vessel patency intra-operatively and post-operatively at the anastomotic site. Baxter Healthcare.

Product Classification:

Class II

Date Initiated: July 11, 2014
Date Posted: August 6, 2014
Recall Number: Z-2126-2014
Event ID: 68789
Reason for Recall:

Baxter Healthcare is recalling the FlowCOUPLER Device because there is a possibility that some units may not maintain electrical continuity during handling intra-operatively and post-operatively, resulting in a loss of Doppler signal. This may result in the inability to monitor anastomotic patency, with the possibility that blood flow may not be appropriately monitored.

Status: Terminated
Product Quantity: 1,315 units
Code Information:

Lot Number SPSGT14-02B0027a SPSGT14-04D0005 SPSGT14-04D0023 SPSGT13-10J0016 SPSGT13-11K0041 SPSGT13-12L0008 SPSGT14-01A0009 SPSGT14-02B0026 SPSGT14-04D0015 SPSGT14-04D0026 SPSGT13-10J0017 SPSGT13-11K0039 SPSGT13-11K0042 SPSGT13-12L0015 SPSGT14-01A0028 SPSGT14-02B0028 SPSGT14-02B0033 SPSGT14-04D0014 SPSGT14-04D0027 SPSGT14-05E0005 SPSGT13-10J0018 SPSGT13-11K0040 SPSGT13-12L0007 SPSGT13-12L0012 SPSGT14-01A0027 Part No. 5151-01200-010 5151-01200-010 5151-01200-010 5151-01200-011 5151-01200-011 5151-01200-011 5151-01200-011 5151-01250-010 5151-01250-010 5151-01250-010 5151-01250-011 5151-01250-011 5151-01250-011 5151-01250-011 5151-01250-011 5151-01300-010 5151-01300-010 5151-01300-010 5151-01300-010 5151-01300-010 5151-01300-011 5151-01300-011 5151-01300-011 5151-01300-011 5151-01300-011

Distribution Pattern:

Worldwide Distribution - USA (nationwide) and the countries of United Kingdom, Germany, Sweden, Spain, Israel and Denmark.

Voluntary or Mandated:

Voluntary: Firm initiated