Baxter Healthcare Corporation: Medical Device Recall in 2022 - (Recall #: Z-0175-2023)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2022.

Data Source: FDA.
Product Description:

Allen Advance Table, RSP-ADVANCE TABLE: a) REF A-71100-US, #107003, b) A-71101-EU, #107003. operating room table

Product Classification:

Class II

Date Initiated: September 20, 2022
Date Posted: November 9, 2022
Recall Number: Z-0175-2023
Event ID: 90848
Reason for Recall:

the adhesive, applied to screws and bolts as an additional holding force, may not cure correctly, and as a result, could result in a loss of holding and reduced mechanical integrity.

Status: Ongoing
Product Quantity: 2 units
Code Information:

a) UDI/DI 00615521003012, Serial Number 122190000035, Lot Number 122220000528 b) UDI/DI 00615521003036, Serial Number 122230000655, Lot Number 122180001460

Distribution Pattern:

MI, BA, WI, WA, TX, and Germany, Japan, South Korea, Sweden, South Africa

Voluntary or Mandated:

Voluntary: Firm initiated