Baxter Healthcare Corporation: Medical Device Recall in 2023 - (Recall #: Z-2208-2023)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2023.

Data Source: FDA.
Product Description:

GEM FLOW COUPLER Monitor, PN 5156-00000-011

Product Classification:

Class II

Date Initiated: June 19, 2023
Date Posted: July 26, 2023
Recall Number: Z-2208-2023
Event ID: 92525
Reason for Recall:

An issue was identified where the LCD monitor display may not indicate the correct active channel. The issue occurs when the monitor is turned on using battery power, and channel B was selected as the last active channel prior to powering down.

Status: Ongoing
Product Quantity: 122 units
Code Information:

GTIN 00844735006148, all serial numbers

Distribution Pattern:

US

Voluntary or Mandated:

Voluntary: Firm initiated