Baxter Healthcare Corporation: Medical Device Recall in 2024 - (Recall #: Z-1329-2024)
See the recall detail below. You can also see other recalls from the same firm in 2024.
HDS Traction Boot II, REF AR-6529-23, The Traction Boot is designed to position, support and/or distract the patient's foot for hip surgery
Class II
Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for the Hip Distraction System, Traction Boot II, due to the boots being assembled incorrectly. The locking ridge on the front face of the boot key socket is facing upwards, which prevents the socket mechanism on the traction boot from attaching properly to the spar of the traction system. This will cause the boot to detach from the traction system.
123510000293, 123510000294, 123510000295, 123510000296, 123510000298, 123510000299, 123510000300, 123510000764, 123510000765, 123510000766, 123510000767, 123510000768, 123510000769, 123510000770, 123510000771, 123510000772, 123510000773, 123510000800, 123510000801, 123510000802, 123510000803, 123510000804, 123510000805, 123510000806, 123510000807, 123510000808, 123510000809, 123510000297, 123350001033, 123350001034, 123350001035, 123350001036, 123350001040, 123350001041, 123350001042, 123360000294, 123360000295, 123360000297, 123360000298, 123360000299, 123360000300, 123360000301, 123360000302, 123360000303, 123360000304, 123360000733, 123360000734, 123360000735, 123360000736, 123360000737, 123360000738, 123360000739, 123360000740, 123360001972, 123360001973, 123360001974, 123360001975, 123360001976, 123360001977, 123360001978, 123360001979, 123360001980, 123360001981, 123350000968, 123350000969, 123350000970, 123350000971, 123350000972, 123350000973, 123350000974, 123350000975, 123350000987, 123350000988, 123350000989, 123350000990, 123350000991, 123350000992, 123350000993, 123350000994, 123350001032
US Nationwide distribution in the state of Florida.
Voluntary: Firm initiated