Baxter Healthcare Corporation: Medical Device Recall in 2024 - (Recall #: Z-1329-2024)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

HDS Traction Boot II, REF AR-6529-23, The Traction Boot is designed to position, support and/or distract the patient's foot for hip surgery

Product Classification:

Class II

Date Initiated: March 11, 2024
Date Posted: March 27, 2024
Recall Number: Z-1329-2024
Event ID: 94202
Reason for Recall:

Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for the Hip Distraction System, Traction Boot II, due to the boots being assembled incorrectly. The locking ridge on the front face of the boot key socket is facing upwards, which prevents the socket mechanism on the traction boot from attaching properly to the spar of the traction system. This will cause the boot to detach from the traction system.

Status: Ongoing
Product Quantity: 80 units
Code Information:

123510000293, 123510000294, 123510000295, 123510000296, 123510000298, 123510000299, 123510000300, 123510000764, 123510000765, 123510000766, 123510000767, 123510000768, 123510000769, 123510000770, 123510000771, 123510000772, 123510000773, 123510000800, 123510000801, 123510000802, 123510000803, 123510000804, 123510000805, 123510000806, 123510000807, 123510000808, 123510000809, 123510000297, 123350001033, 123350001034, 123350001035, 123350001036, 123350001040, 123350001041, 123350001042, 123360000294, 123360000295, 123360000297, 123360000298, 123360000299, 123360000300, 123360000301, 123360000302, 123360000303, 123360000304, 123360000733, 123360000734, 123360000735, 123360000736, 123360000737, 123360000738, 123360000739, 123360000740, 123360001972, 123360001973, 123360001974, 123360001975, 123360001976, 123360001977, 123360001978, 123360001979, 123360001980, 123360001981, 123350000968, 123350000969, 123350000970, 123350000971, 123350000972, 123350000973, 123350000974, 123350000975, 123350000987, 123350000988, 123350000989, 123350000990, 123350000991, 123350000992, 123350000993, 123350000994, 123350001032

Distribution Pattern:

US Nationwide distribution in the state of Florida.

Voluntary or Mandated:

Voluntary: Firm initiated