Baxter Healthcare Corporation: Medical Device Recall in 2025 - (Recall #: Z-1559-2025)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2025.

Data Source: FDA.
Product Description:

Welch Allyn CONNEX Accessory Power Management Stand:

Product Classification:

Class II

Date Initiated: March 24, 2025
Date Posted: April 23, 2025
Recall Number: Z-1559-2025
Event ID: 96478
Reason for Recall:

There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter.

Status: Ongoing
Product Quantity: 114,202 units
Code Information:

All Serial Numbers. Product Code: 7000-APM, UDI-DI: 00732094210613; Product Code: 75-HCA-CTB, UDI-DI: 00732094240597; Product Code: 75-HCA-MTB, UDI-DI: 00732094240603;

Distribution Pattern:

Worldwide - US Nationwide distribution.

Voluntary or Mandated:

Voluntary: Firm initiated