Baxter Healthcare Corporation: Medical Device Recall in 2025 - (Recall #: Z-1962-2025)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2025.
Data Source: FDA.
Product Description:
2.5 V replacement rechargeable batteries for Hillrom WELCH ALLYN PocketScope handles, Product Code 72600
Product Classification:
Class II
Date Initiated: May 9, 2025
Date Posted: June 18, 2025
Recall Number: Z-1962-2025
Event ID: 96825
Reason for Recall:
The replacement rechargeable batteries inserted into the Welch Allyn PocketScope handles have the potential to cause a short circuit in the handle, which can lead to exposure to excessive heat and user burns.
Status: Ongoing
Product Quantity: 277 units
Code Information:
UDI/DI 00732094139488, Lot Number 25-043, Exp. August 2027
Distribution Pattern:
US Nationwide distribution.
Voluntary or Mandated:
Voluntary: Firm initiated