Bayer Medical Care, Inc.: Medical Device Recall in 2018 - (Recall #: Z-2962-2018)
See the recall detail below. You can also see other recalls from the same firm in 2018.
Overhead Counterpoise Systems (OCS), OCS, 580mm CEILING MOUNT, PORTEGRA2, Catalog # OCS CEIL 58, Material Number 84433765
Class II
The torque wrench used on the ceiling column assembly during a recent service visit was outside of the torque specification, potentially overtightening the bolts. The overtightening of bolts may compromise the structural integrity of the system which could lead to the OCS disengaging from the ceiling mount and this may cause injury to the patient or the operator.
Serial Numbers: 917/21956, 516/21035, 817/21916, 817/21910
The devices were distributed to the following US states: OH, TX, and WV.
Voluntary: Firm initiated