Product Description:

WET-FIELD OSHER THERMODOT MARKER, 25GA FINE TIP, CURVED; Part number 221270. Designed to create an ink-free, fine dot on the eye that can be used as a precise reference point for a variety of procedures.

Product Classification:

Class II

Date Initiated: February 26, 2018
Date Posted: May 9, 2018
Recall Number: Z-1539-2018
Event ID: 79908
Reason for Recall:

Incorrect tip was used during manufacturing of the Osher Thermodot Marker. The markers have a needle tip instead of the proper flat tip.

Status: Terminated
Product Quantity: 110 total products
Code Information:

Lot numbers: 6000292 exp. 7/25/20 and 6000540 exp. 08/30/20 UDI: 30886158014899

Distribution Pattern:

Distributed to the states of OH, NY, FL, MN, and KS.

Voluntary or Mandated:

Voluntary: Firm initiated