Beaver-Visitec International, Inc.: Medical Device Recall in 2026 - (Recall #: Z-1008-2026)
See the recall detail below. You can also see other recalls from the same firm in 2026.
BVI Leos Laser and Endoscopy System. Model Number: OME6000U.
Class II
Manufacturing deviation in which the ferrites specified for the USB cable assembly linking the Nurse Monitor touchscreen to the System Controller were not installed as intended. The omission of the ferrites on the USB communication cable results in non-conformance to the design specifications, which require two ferrites (one at each end of the cable).
Model Number: OME6000U. UDI: (01)00886158037576(11)250513(21). Serial Numbers: 77UBV500003, 77UBV500004, 77UBV500005, 77UBV500006, 77UBV500010.
US Nationwide distribution in the states of FL, MI, OK, TX, WI.
Voluntary: Firm initiated