Beckman Coulter Inc.: Medical Device Recall in 2017 - (Recall #: Z-2079-2017)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2017.

Data Source: FDA.
Product Description:

Uric Acid Reagent (URIC 2 x 300), Catalog No. 442785. Intended for the quantitative determination of uric acid concentration in human serum, plasma or urine.

Product Classification:

Class II

Date Initiated: April 3, 2017
Date Posted: May 24, 2017
Recall Number: Z-2079-2017
Event ID: 76962
Reason for Recall:

Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone), Sulfasalazine, and Sulfapyridine and may result in false low results.

Status: Terminated
Product Quantity: 100,286 units total (20,732 units in US)
Code Information:

all lots

Distribution Pattern:

Worldwide Distribution -- Australia, Bangladesh, Belgium, Brunei, Bulgaria, Canada, China, Colombia, Czech Republic, Egypt, Eritrea, Finland, France, Germany, Ghana, Greece, Guatemala, Hong Kong, India, Ireland, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Macau, Malaysia, Mayotte, Mexico, Netherlands, New Zealand, Oman, Pakistan, Panama, Paraguay, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, United States including Puerto Rico, and Vietnam.

Voluntary or Mandated:

Voluntary: Firm initiated