Beckman Coulter Inc.: Medical Device Recall in 2017 - (Recall #: Z-3135-2017)
See the recall detail below. You can also see other recalls from the same firm in 2017.
AQUIOS CL Flow Cytometer System The AQUIOS CL Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping.
Class II
Beckman Coulter has identified that an incorrect setting using a proprietary software configuration tool can result in the Sample Prep Probe of the AQUIOS CL drawing aspirant from the wrong tube in a cassette which manifests in incorrect values being reported out of the laboratory for some patient samples.
All software versions.; Serial numbers - BA30048, AY20032, BA01007, AZ19046, AZ12030, BA25046, BA05011, AY51072, BA01002, AZ15044, BA30049, BA30050, BA05015, AZ19047, BA18029, AY47071, AZ15038, BA23036, BA23035, AZ23058, AZ20057, AY47063, BA15024, BA15023, AZ45080, BA01010, AZ25070, AZ32071, AZ12033, AZ25065, AZ25068, AY51075, AZ19052, BA15025
Worldwide Distribution: United States (nationwide) and countries of: Angola, Australia, Botswana, Canada, Chile, Columbia, Ghana, Israel, Malaysia, Namibia, Pakistan, Panama, South Africa, Uganda, France, Switzerland, Poland, Portugal, Germany, Belgium, Finland, Czech Republic, Denmark, Netherland, Turkey, Spain, Great Britain, Romania, Brazil, Italy, Sweden, Ghana, Colombia, Lesotho and Zambia
Voluntary: Firm initiated