Beckman Coulter, Inc.: Medical Device Recall in 2024 - (Recall #: Z-2545-2024)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

Discrete photometric chemistry analyzer for clinical use.

Product Classification:

Class II

Date Initiated: June 14, 2024
Date Posted: August 21, 2024
Recall Number: Z-2545-2024
Event ID: 94902
Reason for Recall:

Beckman Coulter, Inc. has identified a rise in reagent carousel motion errors in Access 2 Immunoassay System Analyzers that appear to be caused by misaligned index wheels housed within the reagent carousel module. Customers affected by this issue may observe an increase in reagent carousel motion errors or QSMC errors that occur while unloading or loading reagent packs during system initialization and sample processing, which causes a non-operational system status and cancels tests; and patient results may be delayed if these errors cancel test processing.

Status: Ongoing
Product Quantity: 59 units
Code Information:

Catalog Number: 81600N, C11907; UDI: 150995901; Serial Numbers: 574415, 574440, 574484,574416, 574441, 574485, 574417, 574442, 574486, 574418, 574443, 574488,574419, 574444, 574489,574420, 574445, 574490, 574421, 574446, 574491, 574422, 574447 574494,574423, 574448, 574497,574424, 574454, 574502, 574425, 574455, 574503, 574426, 574456 574504,574427, 574457, 574505,574428, 574459, 574506, 574429, 574469, 574507, 574430, 574470 574509,574432, 574471, 574513, 574433, 574472, 574514, 574434, 574477, 574515, 574435, 574479

Distribution Pattern:

Worldwide - US Nationwide distribution in the states of NE, MT, LA, TX, NY, CA, MI, AL, IL the countries of Austria, Belgium, Georgia, India, Indonesia, Italy, Korea, Malaysia, Myanmar, Nigeria, Switzerland, Vietnam.

Voluntary or Mandated:

Voluntary: Firm initiated