Beckman Coulter, Inc.: Medical Device Recall in 2025 - (Recall #: Z-0803-2025)
See the recall detail below. You can also see other recalls from the same firm in 2025.
Access Intact PTH assay, a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of intact parathyroid hormone (parathyrin, PTH) levels in human serum and plasma using the Access Immunoassay Systems, Catalog Number A16972
Class II
Beckman Coulter has identified that some Access PTH reagent packs from lot 339071 may be over- or under-filled in well 0. The dispensed material in well 0 could be above or below specification, risking sealing issues, contamination, instrument errors, delays, or inaccurate results with a negative bias and increase in imprecision. The issue was confirmed on 10 April 2024 by Beckman Coulter through an internal non-conformance (NC-INT-66585). The well 0 scale on the filling line began malfunctioning on the fill immediately before PTH. The malfunction was not noticed until the majority of PTH was filled.
UDI: (01)15099590201937(17)241130(11)231201; Lot 339071
Worldwide distribution - US Nationwide and the country of Canada.
Voluntary: Firm initiated